Thickness Logging with the AlphaGage+

The AlphaGage+ CPDL measures the wall thickness of a pipe or vessel and hence can be used to identify and measure corrosion and/or erosion of the material.

Wear rates can be calculated and used to estimate when the minimum allowable thickness will remain. The instrument can be used for mapping of pipes and vessels with the result held in a table (T-log) for storage, analysis and reporting.

Fig. 1 - Alphagage


Recommended Tool Package

CategoryPart#Description
Acquisition UnitALPHAGAGE+ CPDLCorrosion gauge with accurate thickness data logging
ProbePDEM3750Dual element compression transducer Frequency: 5 MHz Diameter: 3/8"
Calibration Block152062 OR 152060Twin transducer cable TPC-LD OR Twin transducer cable TPC-BD

Please contact our Applications Team if you have any questions. You can also find more solutions on our website.

To follow further content and our newsletter, please contact us and check “Add to mailing list”.